Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200 Quantity: 126 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195200; UDI-DI: 08714729839316; Batch Numbers: 37942457, 37999053, 38224527, 38304916, 38415911, 38687159, 38707977, 38916519, 39029416, 39029419, 39029640, 39092278, 40476286, 40479386, 40523958, 40525677;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210 Quantity: 2547 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195210; UDI-DI: 08714729839323; Batch Numbers: 37914435, 37914436, 37956674, 37981343, 37981344, 38022632, 38022633, 38022634, 38149673, 38181211, 38224528, 38259475, 38259476, 38290713, 38290714, 38290715, 38290716, 38290717, 38399232, 38446223, 38479515, 38506196, 38530446, 38530447, 38629332, 38629333, 38647617, 38647618, 38649501, 38649508, 38687327, 38708440, 38708441, 38745184, 38769513, 38769514, 38934430, 39189795, 39189798, 40466471, 40466472, 40482683, 40482684, 40482685, 40487838, 40487839, 40487900, 40487954, 40487955, 40487956, 40487957, 40501048, 40501049, 40514122, 40514123, 40514222, 40514223, 40533342, 40533344, 40536165, 40536166;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/155; Material Number: M001195220 Quantity: 459 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195220; UDI-DI: 08714729839330; Batch Numbers: 37981345, 38022666, 38629565, 38629566, 38708453, 38745192, 38800928, 38800929, 38916517, 38937262, 38990881, 39074440, 39141261, 40482690, 40482691, 40485783, 40485784, 40489060, 40507284, 40510556, 40512867, 40512868, 40512869, 40512993, 40512994, 40512995, 40513031, 40513032, 40539143, 40539144;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/105; Material Number: M001195230 Quantity: 689 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195230; UDI-DI: 08714729839347; Batch Numbers: 37818205, 37942458, 37956675, 37981346, 38002441, 38224529, 38224530, 38304917, 38304918, 38415912, 38484288, 38506191, 38595896, 38629330, 38647611, 38647613, 38687220, 38687221, 38707976, 38876166, 38970274, 38970936, 39092279, 39092680, 40484896, 40484897, 40484952, 40523918, 40523919, 40525673, 40525674, 40525675, 40525676, 40531394, 40531395, 40531396;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/130; Material Number: M001195240 Quantity: 8542 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195240; UDI-DI: 08714729839354; Batch Numbers: 37818206, 37818207, 37818208, 37818209, 37849382, 37849383, 37849384, 37849385, 37849386, 37849387, 37849388, 37849389, 37914437, 37914438, 37914439, 37914460, 37914461, 37956676, 37956677, 37981347, 38002442, 38022635, 38022636, 38049174, 38112465, 38112466, 38112467, 38112468, 38112469, 38149674, 38149675, 38149676, 38181212, 38181213, 38181214, 38224531, 38224532, 38224533, 38224534, 38259477, 38259478, 38259479, 38290718, 38290719, 38290720, 38290721, 38290722, 38290723, 38350139, 38350320, 38350321, 38350322, 38350323, 38361992, 38361993, 38361994, 38361995, 38361996, 38399233, 38399234, 38399235, 38399236, 38399237, 38399238, 38479516, 38479517, 38506194, 38530448, 38530449, 38560743, 38560744, 38560745, 38560746, 38560747, 38595897, 38595898, 38595899, 38596320, 38596321, 38629336, 38629339, 38629560, 38649503, 38649505, 38687328, 38687329, 38687330, 38687331, 38701165, 38708407, 38708408, 38708411, 38708414, 38708417, 38708418, 38708419, 38745186, 38745187, 38745189, 38769518, 38769519, 38769960, 38769961, 38810595, 38810596, 38810597, 38827709, 38827710, 38827711, 38827712, 38827713, 38827714, 38827715, 38827716, 38827717, 38904801, 38904802, 38904803, 38932705, 38970735, 38970736, 38990278, 38990279, 38990341, 38990343, 38990344, 38990345, 39029646, 39029647, 39029648, 39029649, 39029654, 39074444, 39074445, 39074446, 39074452, 39074453, 39074454, 39074455, 39074456, 39140982, 39140983, 39140984, 39140985, 39140986, 40466475, 40466476, 40466477, 40466585, 40466586, 40466587, 40482558, 40482559, 40482680, 40482681, 40482682, 40488088, 40488089, 40488090, 40488091, 40501038, 40501039, 40501080, 40501081, 40501082, 40501083, 40504691, 40504692, 40507428, 40507429, 40510781, 40510782, 40510783, 40510784, 40510785, 40514124, 40514125, 40521990, 40521991, 40533357, 40533358, 40533359, 40533361, 40533363, 40533365, 40536160, 40536161, 40536162, 40536163, 40536164, 40539613, 40539614, 40539615, 40539616;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/155; Material Number: M001195250 Quantity: 1292 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195250; UDI-DI: 08714729839361; Batch Numbers: 37886785, 37981348, 38022667, 38022668, 38049175, 38149678, 38224535, 38259480, 38259481, 38323413, 38446235, 38506483, 38560459, 38560740, 38560741, 38595642, 38595643, 38644774, 38708452, 38745193, 38810251, 38904563, 38904564, 38937261, 38990880, 39027325, 39027326, 39092681, 39092682, 39140994, 39140995, 39141262, 39141263, 40482688, 40482689, 40502253, 40507285, 40512991, 40512992, 40525712, 40531379, 40537145;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260 Quantity: 358 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195260; UDI-DI: 08714729839378; Batch Numbers: 37818211, 37881104, 37942459, 38002443, 38224536, 38415913, 38595894, 38595895, 38629331, 38687224, 38904567, 38904568, 38916620, 38970937, 38990576, 38990842, 39027322;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/130; Material Number: M001195270 Quantity: 1912 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195270; UDI-DI: 08714729839385; Batch Numbers: 37914462, 37956678, 38079823, 38079827, 38079828, 38181215, 38224537, 38290724, 38399239, 38399240, 38399241, 38446224, 38460466, 38479514, 38479518, 38506192, 38506193, 38527327, 38527328, 38560748, 38560749, 38560750, 38560751, 38629561, 38647619, 38649500, 38649513, 38769494, 38769500, 38827706, 38990346, 38990347, 39074447, 39191592, 40476394, 40482557, 40487835, 40487836, 40507432, 40513948, 40521921, 40521922, 40524021, 40524022, 40533341, 40535977, 40539682;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280 Quantity: 381 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195280; UDI-DI: 08714729839392; Batch Numbers: 38149679, 38323401, 38444072, 38506482, 38506491, 38560458, 38595645, 38810252, 38936816, 39140997, 39141264, 39189785, 39189791, 40482687, 40493382, 40512909;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290 Quantity: 42 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195290; UDI-DI: 08714729839408; Batch Numbers: 38362172, 38362173, 38530450, 38595659;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300 Quantity: 3284 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195300; UDI-DI: 08714729839415; Batch Numbers: 37818213, 37818214, 37818215, 37818216, 37881115, 37886786, 37886787, 37886788, 37914463, 37914464, 37914465, 37943140, 37956679, 37957740, 37977611, 38002444, 38002445, 38085044, 38089225, 38092404, 38092405, 38092406, 38112470, 38112471, 38112472, 38149820, 38181217, 38191752, 38220950, 38330637, 38330638, 38415914, 38415915, 38444073, 38444074, 38484290, 38527320, 38527321, 38560452, 38573062, 38573063, 38585014, 38596980, 38596981, 38629571, 38629572, 38769972, 38769973, 38827700, 38827701, 38827702, 38827703, 38827704, 38827705, 38904809, 38904810, 38916514, 38916515, 38937265, 38990849, 38990852, 38990856, 38990857, 39029815, 39143228, 39143230, 40466130, 40466131, 40466631, 40471499, 40471940, 40476290, 40476291, 40476292, 40479473, 40479474, 40479716, 40479717, 40492475, 40493319, 40493333, 40493440, 40497672, 40499680, 40499681, 40505968, 40505969, 40512591, 40512592, 40512593, 40513026, 40513027, 40521923, 40521924, 40525798, 40525799, 40525860, 40531461, 40531462, 40533302, 40535971, 40535972;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/155; Material Number: M001195310 Quantity: 486 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195310; UDI-DI: 08714729839422; Batch Numbers: 37818217, 37818218, 37818219, 37881110, 37881111, 37881112, 37886789, 37886790, 37886791, 37886792, 37928860, 37928862, 38585015, 38585016, 38745481, 38745482, 38745483, 39143525, 39143526, 39143527, 40466594, 40466595, 40466596, 40476306, 40476307, 40476308, 40482692, 40482693, 40482694, 40485789, 40485790, 40485791;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320 Quantity: 632 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195320; UDI-DI: 08714729839439; Batch Numbers: 37818440, 37818441, 37981349, 38002446, 38200316, 38200317, 38224539, 38304087, 38330639, 38399242, 38399243, 38503335, 38530451, 38629573, 38629576, 38650063, 38650068, 38810584, 38904811, 40466632, 40466633, 40466634, 40476310, 40476442, 40476443, 40482607, 40482609, 40489044, 40491252, 40493253, 40493444, 40502171, 40502326, 40502327, 40504735, 40504736, 40504737, 40525535, 40531443, 40531445, 40533244, 40533305;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/155; Material Number: M001195330 Quantity: 206 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195330; UDI-DI: 08714729839446; Batch Numbers: 37957742, 38361997, 38361998, 38439368, 38527323, 38527324, 38527326, 38876179, 38916511, 39029983, 39029986, 39092697;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/130; Material Number: M001195340 Quantity: 748 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195340; UDI-DI: 08714729839453; Batch Numbers: 37818442, 38181218, 38181219, 38224580, 38240231, 38350326, 38361999, 38460468, 38484291, 38503334, 38530454, 38530455, 38629574, 38650064, 38745484, 38769974, 38769975, 38904812, 38904813, 38937271, 38937272, 38990884, 39092686, 39092687, 40476309, 40476444, 40482608, 40507288, 40510453, 40525536, 40525537;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/155; Material Number: M001195350 Quantity: 384 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195350; UDI-DI: 08714729839460; Batch Numbers: 37881102, 37881103, 37943142, 37943143, 37957743, 38439681, 38439682, 38484292, 38527325, 38876320, 38904816, 38904817, 38916510, 38916512, 38990272, 38990273, 38990274, 39029898, 39092695, 39092696;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400 Quantity: 39 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195400; UDI-DI: 08714729839477; Batch Numbers: 38049178, 38415916, 38595648, 38708403, 39141268;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/130; Material Number: M001195410 Quantity: 902 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195410; UDI-DI: 08714729839484; Batch Numbers: 38061296, 38092407, 38092408, 38112473, 38149821, 38181960, 38198831, 38224581, 38259482, 38273746, 38362200, 38362201, 38399244, 38445017, 38649517, 38649853, 38667794, 38810594, 38904807, 40470780, 40476390, 40482606, 40493383, 40495046, 40495047, 40495048, 40495049, 40502037, 40502268, 40510553, 40510554, 40512596, 40512597, 40512990, 40516627, 40518479, 40518820, 40518821, 40525714, 40525715, 40525716;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/155; Material Number: M001195420 Quantity: 228 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195420; UDI-DI: 08714729839491; Batch Numbers: 38049561, 38065517, 38092409, 38745199, 38745480, 39092274, 40482626, 40482629, 40497899, 40497980, 40502330, 40504597, 40504598, 40531426;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430 Quantity: 76 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195430; UDI-DI: 08714729839507; Batch Numbers: 38149822, 38210076, 38415917, 38595649, 38708406, 39141265;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440 Quantity: 620 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195440; UDI-DI: 08714729839514; Batch Numbers: 38181961, 38198832, 38224582, 38259483, 38324070, 38460469, 38560455, 38560456, 38595654, 38649857, 38745194, 38745195, 38810586, 38810591, 38904805, 39092267, 39092272, 39143222, 40470259, 40478976, 40478977, 40492318, 40497671, 40497858, 40502038, 40502039;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/155; Material Number: M001195450 Quantity: 357 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195450; UDI-DI: 08714729839521; Batch Numbers: 38224583, 38273637, 38649877, 38707975, 38745198, 40470668, 40470669, 40476288, 40476289, 40476304, 40476305, 40495080, 40504684, 40512595, 40516625, 40516626, 40523967, 40525596, 40531425;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470 Quantity: 382 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195470; UDI-DI: 08714729839545; Batch Numbers: 38112474, 38149823, 38181962, 38181963, 38399282, 38415919, 38416360, 38445015, 38649840, 38649849, 39143221, 40470762, 40470763, 40470782, 40476387, 40476388, 40478984, 40478985, 40502172;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/155; Material Number: M001195480 Quantity: 45 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195480; UDI-DI: 08714729839552; Batch Numbers: 38649876, 38668791, 39074172, 39074173, 40482627, 40482628;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/130; Material Number: M001195500 Quantity: 1349 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195500; UDI-DI: 08714729839576; Batch Numbers: 38061298, 38061299, 38092411, 38092412, 38092413, 38112475, 38112476, 38112477, 38167702, 38167705, 38181964, 38181965, 38181966, 38181967, 38181968, 38416361, 38416362, 38595650, 38595651, 38595652, 38650001, 38650003, 38746409, 38746410, 38810247, 38810248, 38904814, 38904815, 38937266, 38937267, 38937268, 38937270, 38990858, 39027320, 39092275, 39092276, 39092277, 40466592, 40466593, 40471963, 40471964, 40478978, 40478979, 40502334, 40502335, 40504653, 40504654, 40513028, 40513029, 40513030, 40524000;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/155; Material Number: M001195510 Quantity: 112 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195510; UDI-DI: 08714729839583; Batch Numbers: 38061601, 38063600, 38434622, 38434623, 38650006, 38650007, 40482630, 40482631, 40502273, 40502274;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/130; Material Number: M001195610 Quantity: 1212 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195610; UDI-DI: 08714729839606; Batch Numbers: 37775581, 37914469, 37914471, 38112478, 38149824, 38149825, 38181969, 38290725, 38342768, 38342769, 38647607, 38647608, 38647615, 38647616, 38649502, 38687333, 38687334, 38708446, 39029659, 40466126, 40466127, 40466473, 40476395, 40485647, 40485732, 40501047, 40504656, 40504657, 40508315, 40512996, 40512997, 40519832, 40519833;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620 Quantity: 598 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195620; UDI-DI: 08714729839613; Batch Numbers: 37775591, 37886793, 37914472, 37914473, 37957744, 38112479, 38149826, 38167704, 38175797, 38444076, 38446234, 38530457, 38745190, 38990885, 39140998, 39140999, 40470358, 40499682, 40499683, 40502254, 40504732, 40504733, 40507436, 40524004, 40539147, 40539467;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/130; Material Number: M001195640 Quantity: 3145 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195640; UDI-DI: 08714729839637; Batch Numbers: 37849390, 37914474, 37914475, 37957745, 38022639, 38022640, 38112640, 38181970, 38181971, 38224584, 38259484, 38290726, 38290727, 38290728, 38399245, 38399246, 38399247, 38444840, 38444841, 38444842, 38460462, 38479512, 38484293, 38484294, 38530458, 38530459, 38560742, 38560752, 38708442, 38708443, 38708444, 38708445, 38769962, 38769963, 38769964, 38810598, 38810599, 38810760, 38810761, 38876169, 38916621, 38970725, 38970726, 38970727, 38970728, 38970732, 38970733, 38970734, 39029800, 39074448, 40466474, 40466590, 40466591, 40476391, 40476392, 40476393, 40485646, 40485726, 40485727, 40485728, 40501045, 40501046, 40504841, 40504842, 40507430, 40507431, 40508310, 40508311, 40512998, 40512999, 40513080, 40523938, 40523939, 40524020, 40533345, 40535978;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650 Quantity: 860 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195650; UDI-DI: 08714729839644; Batch Numbers: 37886794, 37957747, 38022670, 38049176, 38149828, 38175798, 38224585, 38259485, 38323409, 38506199, 38531120, 38560457, 38595646, 38595647, 38745191, 38769968, 38769971, 38916516, 38970729, 38990859, 38990882, 39027327, 39141260, 39189845, 40466132, 40466133, 40466134, 40485781, 40493314, 40502331, 40502332, 40502333, 40504594, 40507434, 40507435, 40531377, 40535710, 40535711, 40537146;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/2RO/130; Material Number: M001195660 Quantity: 156 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195660; UDI-DI: 08714729839651; Batch Numbers: 37914476, 38629575, 38650015, 38687321, 39029882, 39029883, 40466020, 40466021, 40466135, 40507481, 40510684, 40539451;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/2RO/130; Material Number: M001195670 Quantity: 1442 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195670; UDI-DI: 08714729839668; Batch Numbers: 37881100, 37889829, 37914477, 37914478, 37914479, 37943144, 37981351, 37981352, 38002448, 38002449, 38002450, 38017051, 38017052, 38022674, 38022676, 38073300, 38073301, 38273216, 38330740, 38350327, 38362202, 38399248, 38399249, 38434624, 38434625, 38439685, 38439687, 38445013, 38560448, 38560451, 38687322, 38810583, 38810585, 38827798, 38827799, 38902297, 38902298, 38916513, 38990275, 38990276, 38990277, 39092688, 39092689, 39143520, 39189848, 40466136, 40466717, 40476440, 40476441, 40492317, 40492357, 40493443, 40507482, 40510454, 40510455, 40510683, 40510769, 40525862, 40531441, 40531444;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/130; Material Number: M001195710 Quantity: 609 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195710; UDI-DI: 08714729839682; Batch Numbers: 38092414, 38149829, 38181972, 38259486, 38324071, 38416363, 38593058, 38595655, 38629567, 38649848, 38745196, 38810589, 38810590, 38937264, 39092268, 40465990, 40466018, 40466019, 40466137;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/155; Material Number: M001195720 Quantity: 378 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195720; UDI-DI: 08714729839699; Batch Numbers: 38049560, 38092415, 38112641, 38112642, 38181973, 38210072, 38224586, 38224587, 38239068, 38399299, 38595657, 38595658, 38629568, 38650000, 40493316, 40493317, 40495081, 40504596, 40504689, 40504690, 40512594, 40523965, 40525821;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/130; Material Number: M001195740 Quantity: 1135 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195740; UDI-DI: 08714729839712; Batch Numbers: 38092416, 38112643, 38149830, 38181974, 38224588, 38224589, 38259488, 38273747, 38273748, 38290697, 38330742, 38399250, 38416365, 38445016, 38484295, 38593059, 38595656, 38649843, 38649864, 38745197, 38810587, 38810592, 38876171, 38876172, 38904806, 38937263, 38990883, 39092270, 39143220, 39189853, 39189855, 40485616, 40485617, 40492473, 40493384, 40497670, 40497857, 40499760, 40499761, 40502300, 40502325, 40507454, 40518477, 40518478, 40525595, 40525718, 40525719, 40536127;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/155; Material Number: M001195750 Quantity: 432 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195750; UDI-DI: 08714729839729; Batch Numbers: 38049562, 38112644, 38112645, 38181975, 38259489, 38273638, 38342770, 38346223, 38399380, 38707974, 39074174, 39074176, 39092273, 40470667, 40476303, 40485786, 40485787, 40493318, 40495100, 40507439, 40512986, 40512987, 40523966, 40524003, 40531398, 40531399;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/STRAIGHT/1RO/130; Material Number: M001195810 Quantity: 658 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195810; UDI-DI: 08714729839743; Batch Numbers: 37914480, 37957748, 38022641, 38149831, 38224590, 38446222, 38649514, 39074449, 39074450, 39074451, 39187259, 39187840, 39187841, 40476396, 40485725, 40493312, 40504693, 40507433, 40510780, 40513102, 40513103;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/130; Material Number: M001195840 Quantity: 7945 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195840; UDI-DI: 08714729839774; Batch Numbers: 37818448, 37818450, 37849391, 37849392, 37849393, 37849394, 37914481, 37914482, 37914483, 37957749, 37957750, 37957751, 37957752, 37981353, 37981354, 37981355, 37981356, 38002451, 38002452, 38002453, 38002454, 38002455, 38002456, 38022642, 38022643, 38022644, 38022645, 38049173, 38065519, 38065520, 38065522, 38065523, 38065524, 38065525, 38065527, 38065528, 38065529, 38065531, 38065532, 38065533, 38106627, 38106628, 38106630, 38149832, 38149833, 38149834, 38149835, 38149836, 38181976, 38181977, 38181978, 38181979, 38181980, 38259490, 38259491, 38290729, 38290730, 38290731, 38290732, 38290733, 38350328, 38350329, 38350330, 38350331, 38350332, 38350333, 38362203, 38362205, 38362206, 38531121, 38629563, 38649510, 38687335, 38708448, 38708449, 38708450, 38708451, 38769965, 38769966, 38769967, 38827707, 38827708, 38876170, 38904804, 38932706, 38932707, 38934432, 38934433, 38934434, 39029655, 39029656, 39029657, 39029658, 39074458, 39074459, 39074600, 39074601, 39074602, 39074603, 39140987, 39140988, 39140989, 39140990, 39140991, 39140992, 39187842, 39187844, 39187845, 39187846, 39219921, 40466468, 40466469, 40466470, 40466588, 40466589, 40466607, 40466608, 40466609, 40466610, 40471679, 40471960, 40471961, 40476398, 40476480, 40476481, 40476482, 40485645, 40485722, 40485723, 40485724, 40485729, 40485730, 40485731, 40493372, 40493373, 40493374, 40493375, 40493376, 40493377, 40493378, 40493379, 40493380, 40493381, 40494944, 40494945, 40494946, 40502255, 40502258, 40502259, 40504658, 40504659, 40504840, 40508312, 40508313, 40510558, 40510559, 40512915, 40512916, 40512918, 40513100, 40513101, 40521985, 40521986, 40521987, 40521988, 40521989, 40523930, 40523931, 40523932, 40523933, 40523934, 40523935, 40523936, 40523937, 40526446, 40526447, 40526448, 40526449, 40526450, 40526451, 40528687, 40528688, 40528689, 40528690, 40528691, 40533346, 40533347, 40533348, 40533349, 40535973, 40535974, 40535976, 40539619;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850 Quantity: 926 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195850; UDI-DI: 08714729839781; Batch Numbers: 37996606, 38022677, 38022678, 38065534, 38065535, 38112646, 38149837, 38224591, 38259492, 38290688, 38446231, 38769969, 38769970, 38810249, 38875417, 38936818, 38970730, 38970731, 38990886, 39074178, 40466463, 40466605, 40466606, 40470665, 40470666, 40476198, 40484940, 40485711, 40485712, 40485782, 40492387, 40507437, 40512910, 40512911, 40512912, 40512914, 40523982, 40524002, 40525713, 40531378, 40537147;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/027/STRAIGHT/1RO/130; Material Number: M001195910 Quantity: 86 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195910; UDI-DI: 08714729839828; Batch Numbers:38240235, 38273744, 39092271, 40464692, 40464699, 40531397;
Boston Scientific Cork — The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/027/BERN/1RO/130; Material Number: M001195940 Quantity: 550 The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment. Codes: Material Number: M001195940; UDI-DI: 08714729839842; Batch Numbers:38061297, 38092418, 38149838, 38181982, 38224592, 38240234, 38240236, 38273745, 38649863, 38810593, 39143039, 39189852, 40466138, 40466139, 40466464, 40470781, 40476389, 40478986, 40497669, 40502265, 40507286, 40512989, 40525717;